WORLDWIDE

Products

40W-65W Medical Power Supply Open Frame

PW-IM065B2-1YXXXZ

65W AC-DC Open Frame
L 101.6 × W 50.8 × H 2 7.5 mm

Features
*Both ITE & Medical Approvals.
*Medical Body Floating ( BF ) rated, Class I Power Supply
*Medical applications Protection: Means of patient protection ( MOPP )
*Altitude during operation: ITE up to 5000m ,Medical Below 3000m.
*Less than 75mW no-load power consumption
*Low leakage current
*3 year warranty
View More Details
PDF Datasheet (402 KB)
Description
*Dimension : L101.6 × W50.8 × H27.5 mm ( 4" x 2" x 1.08" )
*Weight : 0.136 kgs. ( 0.30 lbs. )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Inrush current : < 45A peak @115VAC; < 90A peak @230VAC cold start @25°C
  • Input current ( rms ) : 2A @115VAC; 1A @230VAC max.
  • Efficiency : > 87% typical @full load, 115VAC ( 5V series excluded )
  • Touch current < 100uA @264VAC
  • Maximum output power ( Po ) : 65 Watts convection cooling
  • Hold-up time : > 10ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over load protection : 110% to 180% maximum rating, Auto-recovery
  • Over voltage protection : Latching type. AC Recycle
Safety standards
  • UL60601-1 3.1rd Edition
  • UL/c-UL UL60950-1
  • UL/c-UL UL62368-1
  • EN60601-1 3.1rd Edition
  • TUV EN60950-1
  • TUV EN62368-1
  • IEC EN60601-1 3.1rd Edition
  • CB IEC 60950-1
  • CB IEC 62368-1
EMC standards
  • IEC60601-1-2 Ed4:2014
  • EN60601-1-2 Ed4:2015
  • EN55024
  • EN 55011 Class B
  • EN55032 Class B
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +70°C ( Derate linearly 2.5% per °C from 51 to 70°C, Derate linearly 1.0% per °C from 0 to -20°C )
  • Storage temperature: -20°C to +85°C
  • Humidity: Non-condensing 10% to 90%
  • MTBF : > 150,000 hours @full load and 25°C ambient temperature
DC OUTPUT & FEATURES
Model Output Voltage Minimun Load Maximun
Load
Po Average efficiency Output
Regulation
Ripple & Noise (Vp-p)
Full Load < 30% Load
PW-IM040B2-1Y050Z +5V 0A 8.00A 40W > 86% ±5% 100mV 200mV
PW-IM065B2-1Y120Z +12V 0A 5.42A 65W > 89% ±3% 120mV 200mV
PW-IM065B2-1Y150Z +15V 0A 4.34A 65W > 89% ±3% 120mV 300mV
PW-IM065B2-1Y190Z +19V 0A 3.43A 65W > 89% ±3% 120mV 300mV
PW-IM065B2-1Y240Z +24V 0A 2.71A 65W > 89% ±3% 200mV 300mV
PW-IM065B2-1Y280Z +28V 0A 2.33A 65W > 89% ±3% 200mV 350mV
PW-IM065B2-1Y480Z +48V 0A 1.36A 65W > 89% ±2% 200mV 480mV
PW-IM065B2-1Y540Z +54V 0A 1.21A 65W > 89% ±2% 200mV 480mV
Note:
  • In/Output connector options: Z=J ( JST VH Type Connector or Equ. ), Z=M(TKP P8800I type).
  • Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
  • The ripple and noise will be higher while load under 30% of rate load during Burst Mode operation.
    The Burst Mode operation is to meet the latest efficiency regulations.
  • The output voltage is set by manufacturer.