Toggle menu

IEC 60601 Power Supply with BF Rating for Healthcare Equipment

The PW-IM065B2-1YXXXZ is a 65W AC-DC open-frame power supply from POWER-WIN TECHNOLOGY CORP., designed for integration into medical and ITE equipment. With a compact 101.6 × 50.8 × 27.5 mm form factor, convection cooling, low leakage current, and less than 75mW no-load power consumption, this IEC 60601 power supply supports space-conscious equipment designs. It provides multiple output voltage options from 5V to 54V, with short-circuit and overload auto-recovery protection.

Designed around medical equipment requirements, the IEC 60601 power supply supports BF-rated applications, Class I power supply construction, and means of patient protection (MOPP). It complies with listed UL, EN, IEC, EMC, CE, and related safety standards, while its operating temperature range of -20°C to +70°C and MTBF above 150,000 hours provide defined operating specifications. As a low leakage current medical power supply, it is positioned for equipment designs where leakage-current requirements, compact installation, and reliable power conversion need to be considered together. The 65W medical power supply provides a defined solution for medical and healthcare equipment integration.
65W AC-DC Open FrameL 101.6 × W 50.8 × H 2 7.5 mm, PW-IM065B2-1YXXXZ

PW-IM065B2-1YXXXZ

65W AC-DC Open Frame
L 101.6 × W 50.8 × H 2 7.5 mm

*Both ITE & Medical Approvals.
*Medical Body Floating ( BF ) rated, Class I Power Supply
*Medical applications Protection: Means of patient protection ( MOPP )
*Altitude during operation: ITE up to 5000m ,Medical Below 3000m.
*Less than 75mW no-load power consumption
*Low leakage current
*3 year warranty
Description :

*Dimension : L101.6 × W50.8 × H27.5 mm ( 4" x 2" x 1.08" )
*Weight : 0.136 kgs. ( 0.30 lbs. )

Specification :

Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Inrush current : < 45A peak @115VAC; < 90A peak @230VAC cold start @25°C
  • Input current ( rms ) : 2A @115VAC; 1A @230VAC max.
  • Efficiency : > 87% typical @full load, 115VAC ( 5V series excluded )
  • Touch current < 100uA @264VAC
  • Maximum output power ( Po ) : 65 Watts convection cooling
  • Hold-up time : > 10ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over load protection : 110% to 180% maximum rating, Auto-recovery
  • Over voltage protection : Latching type. AC Recycle
Safety standards
  • UL60601-1 3.1rd Edition
  • UL/c-UL UL60950-1
  • UL/c-UL UL62368-1
  • EN60601-1 3.1rd Edition
  • TUV EN60950-1
  • TUV EN62368-1
  • IEC EN60601-1 3.1rd Edition
  • CB IEC 60950-1
  • CB IEC 62368-1
EMC standards
  • IEC60601-1-2 Ed4:2014
  • EN60601-1-2 Ed4:2015
  • EN55024
  • EN 55011 Class B
  • EN55032 Class B
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +70°C ( Derate linearly 2.5% per °C from 51 to 70°C, Derate linearly 1.0% per °C from 0 to -20°C )
  • Storage temperature: -20°C to +85°C
  • Humidity: Non-condensing 10% to 90%
  • MTBF : > 150,000 hours @full load and 25°C ambient temperature
DC OUTPUT & FEATURES
Model Output Voltage Minimun Load Maximun
Load
Po Average efficiency Output
Regulation
Ripple & Noise (Vp-p)
Full Load < 30% Load
PW-IM040B2-1Y050Z +5V 0A 8.00A 40W > 86% ±5% 100mV 200mV
PW-IM065B2-1Y120Z +12V 0A 5.42A 65W > 89% ±3% 120mV 200mV
PW-IM065B2-1Y150Z +15V 0A 4.34A 65W > 89% ±3% 120mV 300mV
PW-IM065B2-1Y190Z +19V 0A 3.43A 65W > 89% ±3% 120mV 300mV
PW-IM065B2-1Y240Z +24V 0A 2.71A 65W > 89% ±3% 200mV 300mV
PW-IM065B2-1Y280Z +28V 0A 2.33A 65W > 89% ±3% 200mV 350mV
PW-IM065B2-1Y480Z +48V 0A 1.36A 65W > 89% ±2% 200mV 480mV
PW-IM065B2-1Y540Z +54V 0A 1.21A 65W > 89% ±2% 200mV 480mV
Note:
  • In/Output connector options: Z=J ( JST VH Type Connector or Equ. ), Z=M(TKP P8800I type).
  • Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
  • The ripple and noise will be higher while load under 30% of rate load during Burst Mode operation.
    The Burst Mode operation is to meet the latest efficiency regulations.
  • The output voltage is set by manufacturer.
  • Send Message
    Contact Information

    By submitting your contact information, you acknowledge that you consent to our processing data in accordance with the Privacy and Cookie Policy.

    For equipment designers, the PW-IM065B2-1YXXXZ series provides several output-voltage options, including 5V, 12V, 15V, 19V, 24V, 28V, 48V, and 54V, with output power available at 40W or 65W depending on the model. The open-frame construction uses convection cooling, while short-circuit and overload protection feature auto-recovery. Input/output connector options are also specified, allowing selection according to equipment integration requirements. These details make the IEC 60601 power supply relevant when engineers need to match output voltage, current capacity, physical dimensions, and connector configuration during product development.

    The IEC 60601 power supply is specified for operation from 90–264VAC and 47–63Hz, with an operating temperature range of -20°C to +70°C and MTBF above 150,000 hours at full load and 25°C ambient temperature. Its low leakage current and touch current below 100µA at 264VAC are particularly relevant to medical equipment designs. For projects requiring a compact power source with defined electrical, safety, EMC, and environmental specifications, the 65W medical power supply and low leakage current medical power supply configurations provide clearly documented parameters for equipment selection and integration.