Features
Power-Win is a professional power supply manufacturer located in Taipei, Taiwan. Starting with PC power supplies business in 1989, Power-Win extends its product lines into Desktop Power Supply, open frame power supplies, DC / DC ATX converters, and medical power supplies. With over 20 year experience in design, research, and manufacturing, Power-Win not only develops its own products to lead market trends and also works with customers for OEM / ODM projects. To meet latest energy saving demand and green power trend, we introduce more efficiency level V adapters and also open frame units to fulfill customer’s needs. Etc. PW-M065B-1Yxxx (Medical Power Supply)

*Both Medical 60601-1 3rd Edition MOPP and ITE approved
*Compact, High Power density, 40W/65W in 2" x 3" footprint
*Meet DOE level VI 88% & ErP Tier 1 requirements.(Except +5V mode 86%)
*No load power consumption < 0.21W@230VAC (Except +5V mode ,< 0.1W@230VAC max)
*Ultra low Earth leakage current < 189uA @264VAC
*2 year warranty
PDF Datasheet (577 KB)
Description
*Dimension : L76.2 × W50.8 × H28.7 mm ( 3" × 2" × 1.13" )
*Weight : 0.12 kgs. ( 0.26 lbs. )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Inrush current : < 65A peak @115VAC; < 130A peak @230VAC cold start @25°C
  • Input current (rms) : 1A @115VAC; .0.6A @230VAC max.
  • Efficiency : > 87% typical @full load, 115VAC(Except +5V mode ,85% typical @full loadl)
  • Maximum output power (Po) : 65 Watts convection cooling((Except +5V mode ,40 Watts )
  • Hold-up time : > 10ms typical @ load 75% , 115VAC
  • Short circuit protection : Auto-recovery
  • Over load protection : 105% to 170% maximum rating, Auto-recovery
  • Over voltage protection : Shutdown and latch off, AC recycle
Environmental
  • Operating temperature : -20 to +70°C (Derate linearly 2.5% per °C from 41 to 70°C)
  • Storage temperature: -20°C to +85°C
  • Humidity: Non-condensing 0% to 90%
  • MTBF : > 100,000 hours @full load and 25°C ambient temperature
SAFETY STANDARDS
    Medical
  • 60601-1 3rd Edition
  • CSA-C22.2 No.60601-1 3rd Edition
  • EN60601-1 3rd Edition
  • IEC EN60601-1 3rd Edition
  • ITE
  • UL UL60950-1
  • UL UL62368-1
  • CSA-C22.2 NO.60950-1
  • CSA-C22.2 NO.62368-1
  • TUV EN60950-1
  • TUV EN62368-1
  • CB IEC 60950-1
  • CB IEC 62368-1
EMI STANDARDS
    Medical
  • IEC60601-1-2 Ed4:2014
  • EN60601-1-2 Ed4:2015
  • FCC Part 18 Class B
  • EN 55011 Class B
  • CE
  • ITE
  • EN55032 Class B
  • EN55024
  • CISPR22 Class B
  • FCC Part 15 Class B
  • CE
DC OUTPUT & FEATURES
※ Other DC output 40W for 5V, 65W for 12-56.9V.
Model no. Output
voltage
Minimum
load
Maximum
load convection
Po O/P
Regulation
Ripple & Noise
(Vp-p)
Efficiency Level
PW-IM040B-1Y050 +5V 0A 8A 40W ±5% 150mV VI
PW-IM065B-1Y120 +12V 0A 5.416A 65W ±3% 240mV VI
PW-IM065B-1Y150 +15V 0A 4.333A 65W ±3% 300mV VI
PW-IM065B-1Y190 +19V 0A 3.421A 65W ±3% 300mV VI
PW-IM065B-1Y200 +20V 0A 3.25A 65W ±3% 300mV VI
PW-IM065B-1Y240 +24V 0A 2.708A 65W ±3% 300mV VI
PW-IM065B-1Y280 +28V 0A 2.321A 65W ±3% 300mV VI
PW-IM065B-1Y480 +48V 0A 1.354A 65W ±3% 300mV VI
PW-IM065B-1Y560 +56V 0A 1.161A 65W ±3% 300mV VI
Note:
  • At 25°C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
  • Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 100uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
  • DC OUTPUT & FEATURES Operating temperature is 0 to +40°C.
  • -20 to 0°C output characteristics Output regulation are ±10%.