Products

90W-110W Medical Adapter

PW-IM110A-1YXXXN

110W AC-DC Desktop Adaptor
AC Inlet-C8

Features
*ITE & Medical Dual Safety Approvals
*2-Pole AC inlet IEC320-C8 Class II
*Medical safety approved ( 2x MOPP between primary to secondary )
*Suitable for BF application with appropriate system consideration
*Altitude : ITE / Medical up to 5000m.
*Rectangular plastic case with IP41 level
*< 0.21W No Load Input Power
*Touch current less than 100uA
*3 years warranty
View More Details
PDF Datasheet (505 KB)
Description
Dimension : L140 × W63 × H33 mm ( 5.5" × 2.48" × 1.29")
Weight :0.78 kgs. ( 1.72 lbs. )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Power Factor : > 0.95 @115VAC; > 0.90 @230VAC @full load
  • Inrush current : < 50A peak @115VAC < 120A peak @230VAC cold start @25°C
  • Input current ( rms ) : 2A @90VAC; 1.25A @264VAC max.
  • Efficiency : > 88% @average load, 115/230VAC
  • Hold–up time : > 10ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over Load Protection: Auto-recovery
  • Over Voltage Protection: latch off
Safety standards
  • UL/cUL 60601-1
  • TUV EN 60601-1
  • CB IEC 60601-1
  • UL/cUL UL 62368-1
  • TUV EN 62368-1
  • CB IEC 62368-1
EMC standards
  • EN 60601-1-2
  • EN 55011 Class B
  • EN (BS ES) 55032 Class B
  • EN (BS ES) 55035 Class B
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +60°C (Refer to derating curve )
  • Operating Humidity: 0%~95% , non-condensing.
  • Storage temperature : -20°C to +85°C
  • Storage Humidity: 0%~95% , non-condensing.
  • MTBF : 350,000 hours min. at full load and 25°C ambient temperature(Base on Bellcore TR-332)
Standard output cable
  • AWM 11353#14 1C+Shielding, UL 80°C 300V VW-1, 1000mm.
  • AWM 11353#16 1C+Shielding, UL 80°C 300V VW-1, 1000mm.
  • Depends on customers requirements.
Model Output voltage Output current Po O/P Regulation Ripple & Noise
(Vp-p)
Energy
Level
PW-IM105A-1Y120N 12V 8.75A 105W +/- 5% 150mV VI
PW-IM110A-1Y150N 15V 7.34A 110W +/- 5% 150mV VI
PW-IM110A-1Y190N 19V 5.59A 110W +/- 5% 150mV VI
PW-IM110A-1Y240N 24V 4.49A 110W +/- 3% 150mV VI
PW-IM110A-1Y280N 28V 3.93A 110W +/- 3% 150mV VI
PW-IM110A-1Y360N 36V 3.06A 110W +/- 3% 150mV VI
PW-IM110A-1Y480N 48V 2.29A 110W +/- 3% 150mV VI
Note:
1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 47uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
2. The switching frequency of this series is set within 45 to 65KHz at full load.
3. The ripple and noise of this series is tested under full load condition.

Loading...