WORLDWIDE

Products

40W-65W Medical Adapter
Medical AC/DC Power Adapter (40/60/65Watts Series Level VI and EU COC Tier 2)

PW-M065A2-1YXXXA

65W AC-DC Desktop Adaptor
AC Inlet-C14

Features
*Medical 60601-1 3.1rd Edition approved
*Meet IEC60601-1-2 Ed4:2014
*3-pole IEC320-C14 AC inlet, Class I
*Altitude: 5,000M, 2 x MOPP, BF rated
*US DOE and EU ErP Efficiency Compliant
*No load Power consumption < 150mW
*Earth leakage current : < 100uA
*3 year warranty
View More Details
PDF Datasheet (468 KB)
Description
*Dimension : L132 × W69 × H40 mm ( 5.2" × 2.72" × 1.57" )
*Weight : 0.39 kgs. ( 0.86 lbs )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Inrush current : < 50A peak @115VAC; < 90A(5V 120A) peak @230VAC cold start @25°C
  • Input current (rms) : 2A @90VAC; 1A @264VAC max.
  • Efficiency : > 89% @average Efficiency, 115/230VAC (Except +5V model)
  • Maximum output power (Po) : 40Watts / 60Watts / 65Watts
  • Hold–up time : > 16ms typical @full load, 115VAC
  • Short Circuit Protection: Auto-recovery
  • Over Load Protection: Auto-recovery
  • Over Voltage Protection: Latch off, AC recycle
Safety standards
  • UL/c-UL 60601-1 3.1rd Edition
  • TUV EN60601-1 3.1rd Edition
  • CB IEC EN60601-1 3.1rd Edition
EMC standards
  • EN60601-1-2 Ed4:2015
  • IEC60601-1-2 Ed4:2014
  • EN55011 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +60°C (Derate linearly 2.5% per °C from 51 to 60°C)
  • Storage temperature : -20°C to +85°C
  • Humidity: 20% ~ 90%.
  • MTBF : > 100,000 hours at full load and 25°C ambient temperature
Model Output Rating Po O/P Regulation Ripple & Noise
(Vp-p)
Energy
Level
PW-M040A2-1Y050A +5V 8A 40W ±5% 120mV VI
PW-M060A2-1Y120A +12V 5A 60W ±5% 1200mV VI
PW-M060A2-1Y150A +15V 4A 60W ±5% 150mV VI
PW-M065A2-1Y190A +19V 3.42A 65W ±3% 190mV VI
PW-M065A2-1Y240A +24V 2.7A 65W ±3% 240mV VI
PW-M065A2-1Y280A +28V 2.32A 65W ±3% 280mV VI
PW-M065A2-1Y360A +36V 1.81A 65W ±3% 360mV VI
PW-M065A2-1Y480A +48V 1.35A 65W ±3% 480mV VI
Note:
  • 1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.