WORLDWIDE

Products

130/140W Medical Adapter
  • Facebook
  • LINE
  • Twitter
  • LinkedIn
DOE Level VI & EU ErP Tier 2 Efficiency Compliant,Medical AC/DC Power Adapter AD-A7 V2 Series 130/140Watts

PW-M140A2-1YXXXG

140W AC-DC Desktop Adaptor
AC Inlet-C14

Features
*Medical Approvals
*3-prong IEC 320-C14, Class I
*Medical safety approved ( 2x MOPP between primary to secondary )
Suitable for BF application with appropriate system consideration.
*Altitude during operation: Medical up to 5000m.
*≦ 0.15W No Load Input Power
*Touch current less than 100uA
*Meet IEC60601-1-2 Ed4:2014
*3 year warranty
View More Details
PDF Datasheet (204 KB)
Description
*Dimension :L170 ×W64.8 ×H38.5mm ( 6.69"×2.55"×1.52")
*Weight : 0.61 kgs. ( 1.34 lbs. )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Power Factor : > 0.95 @115VAC; > 0.90 @230VAC @full load
  • Inrush current : < 50A peak @115VAC; < 100A peak @230VAC cold start @25°C
  • Input current (rms) : 2.5A @115VAC; 1.3A @230VAC max.
  • Efficiency : > 89% (Average) 25 %, 50 %, 75 % and 100 % of full load @ 115/230VAC
  • Touch current < 100uA @264VAC
  • Maximum output power (Po) : 140 Watts convection cooling
  • Hold-up time : > 20ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over load protection : 110% to 150% maximum power, Auto-recovery
  • Over voltage protection : Shutdown and latch off, AC recycle
Safety standards
  • UL/cUL 60601-1 3.1rd Edition
  • TUV EN60601-1 3.1rd Edition
  • CB IEC 60601-1 3.1rd Edition
EMC standards
  • EN60601-1-2 Ed4:2015
  • IEC60601-1-2 Ed4:2014
  • EN 55011 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +60°C (Refer to "Derating Curve")
  • Cooling: Convection-cooling
  • Storage temperature: -20°C to +85°C
  • Humidity: Non-condensing 0% to 95%
  • MTBF : > 120,000 hours @full load and 25°C ambient temperature based on MIL-HDBK-217F
Standard output wires
  • AWM 1571 # 14 AWG / 1C + SHIELDING, UL 80°C 30V VW-1, 1050 mm.
  • AWM 1185 # 16 AWG / 1C + SHIELDING, UL 80°C 300V VW-1, 1050 mm.
  • Depends on customers requirements.
DC OUTPUT & FEATURES
Model No. Output Rating Po Output Regulation Ripple & Noise
(Vp-p)
DOE Efficiency EU ErP
PW-M130A2-1Y120G +12V 10.84A 130W ±5% 120mV Level VI Tier 2 
PW-M130A2-1Y150G +15V 8.67A 130W ±5% 150mV Level VI Tier 2 
PW-M140A2-1Y190G +19V 7.37A 140W ±5% 190mV Level VI Tier 2 
PW-M140A2-1Y240G +24V 5.84A 140W ±3% 240mV Level VI Tier 2 
PW-M140A2-1Y280G +28V 5.00A 140W ±3% 280mV Level VI Tier 2 
PW-M140A2-1Y360G +36V 3.89A 140W ±3% 280mV Level VI Tier 2 
PW-M140A2-1Y480G +48V 2.92A 140W ±3% 280mV Level VI Tier 2 
PW-M140A2-1Y540G +54V 2.60A 140W ±3% 280mV Level VI Tier 2 
Note:1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
Derating curve
Loading...