WORLDWIDE

Products

250W Medical Adapter

PW-250A-1YXXXP1MZS

250W AC-DC Desktop Adaptor
AC Inlet-C14, with I/O Switch

Features
*ITE 60950-1, 62368-1 & Medical 60601-1 approvals. ( Class I )
*3-prong IEC320-C14, Class I (I/O Switch Option)
*Medical safety approved ( 2x MOPP between primary to secondary )
Suitable for BF application with appropriate system consideration.
*Altitude during operation: ITE/Medical up to 5000m.
*≦ 0.15W No Load Input Power
*Touch current less than 100uA
*Meet IEC60601-1-2 Ed4:2014
*3 year warranty
View More Details
PDF Datasheet (563 KB)
Description
*Dimension : L 200×W 90×H45 mm ( 7.874"x3.543"x1.77" )
*Weight : 1.3 kgs. ( 2.87 lbs. )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Power Factor : > 0.95 @115VAC; > 0.90 @230VAC @full load
  • Inrush current : < 40A peak @115VAC; < 80A peak @230VAC cold start @25°C
  • Input current (rms) : 2.5A @115VAC; 1.3A @230VAC max.
  • Efficiency : > 89% (Average) 25 %, 50 %, 75 % and 100 % of full load @ 115/230VAC
  • Touch current < 100uA @264VAC
  • Maximum output power (Po) : 250 Watts convection cooling
  • Hold-up time : > 10ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over load protection : 105% to 160% maximum power, Auto-recovery
  • Over voltage protection : Auto-recovery
Safety standards
  • UL/cUL 60601-1 3.1rd Edition
  • TUV EN60601-1 3.1rd Edition
  • CB IEC 60601-1 3.1rd Edition
  • UL/c-UL UL60950-1
  • UL/c-UL UL62368-1
  • TUV EN60950-1
  • TUV EN62368-1
  • CB IEC 60950-1
  • CB IEC 62368-1
EMC standards
  • EN60601-1-2 Ed4:2015
  • IEC60601-1-2 Ed4:2014
  • EN 55011 Class B
  • EN 55032 Class B
  • EN 55024 Class B
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE.
Environmental
  • Operating temperature : -20 to +60°C (Refer to "Derating Curve")
  • Cooling:Convection-cooling
  • Storage temperature : -20°C to +85°C
  • Humidity: Non-condensing 0% ~ 95%.
  • MTBF : > 200,000 hours @full load and 25°C ambient temperature based on Bellcore TR-332
DC OUTPUT & FEATURES
Model Output Rating Po O/P regulation Ripple & Noise
(Vp-p)
DOE Efficiency Eu ErP
PW-250A-1Y120P1MZ +12V 20.83A 250W ±5% 240mV VI tier 2
PW-250A-1Y190P1MZ +19V 13.157A 250W ±5% 300mV VI tier 2
PW-250A-1Y240P1MZ +24V 10.416A 250W ±5% 300mV VI tier 2
PW-250A-1Y280P1MZ +28V 8.928A 250W ±5% 300mV VI tier 2
PW-250A-1Y360P1MZ +36V 6.94A 250W ±5% 300mV VI tier 2
PW-250A-1Y480P1MZ +48V 5.208A 250W ±5% 300mV VI tier 2
PW-250A-1Y540P1MZ +54V 4.629A 250W ±5% 300mV VI tier 2
Note:
1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 47uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
2. AC INLET : IEC 320-C14(3PIN)
3. Z = S to denote With I/O Switch, Z = X to denote Without I/O Switch.
Loading...