AC Inlet-C8

Features
*ITE 60950-1, 62368-1 & Medical 60601-1 approvals.
*2-Pole AC inlet IEC320-C8 Class II
*Medical safety approved ( 2x MOPP between primary to secondary )
*Suitable for BF application with appropriate system consideration.
*Altitude : ITE / Medical up to 5000m.
*≦ 0.15W No Load Input Power
*Touch current less than 100uA
*Rectangular plastic case with IP41 level
*3 years warranty
View More Details
PDF Datasheet (485 KB)
Description
Dimension : L140 × W63 × H33 mm ( 5.5" × 2.48" × 1.29")
Weight : 0.5 kgs. ( 1.10 lbs. )
Specification
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Power Factor : > 0.95 @115VAC; > 0.90 @230VAC @full load
  • Inrush current : < 50A peak @115VAC < 120A peak @230VAC cold start @25°C
  • Input current ( rms ) : 2A @90VAC; 1.25A @264VAC max.
  • Efficiency : > 89% @average load, 115/230VAC
  • Hold–up time : > 10ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over Load Protection: Auto-recovery
  • Over Voltage Protection: latch off
Safety standards
  • UL/c-UL 60601-1 3.1rd Edition
  • TUV EN60601-1 3.1rd Edition
  • CB IEC60601-1 3.1rd Edition
  • UL/c-UL UL60950-1
  • UL/c-UL UL62368-1
  • TUV EN60950-1
  • TUV EN62368-1
  • CB IEC 60950-1
  • CB IEC 62368-1
  • SGS IEC/EN 60601-1-11 2rd Edition
EMC standards
  • EN60601-1-2 Ed4:2015
  • IEC60601-1-2 Ed4:2014
  • EN 55011 Class B
  • EN55032 Class B
  • EN55024 Class B
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +60°C (Refer to derating curve )
  • Operating Humidity: 0%~95% , non-condensing.
  • Storage temperature : -20°C to +85°C
  • Storage Humidity: 0%~95% , non-condensing.
  • MTBF : 350,000 hours min. at full load and 25°C ambient temperature(Base on Bellcore TR-332)
Standard output cable
  • AWM 11353#14 1C+Shielding, UL 80°C 300V VW-1, 1000mm.
  • AWM 11353#16 1C+Shielding, UL 80°C 300V VW-1, 1000mm.
  • Depends on customers requirements.
Model Output voltage Output current Po O/P Regulation Ripple & Noise
(Vp-p)
Energy
Level
PW-IM105A-1Y120N 12V 8.75A 105W +/- 5% 150mV VI
PW-IM110A-1Y150N 15V 7.34A 110W +/- 5% 150mV VI
PW-IM110A-1Y190N 19V 5.59A 110W +/- 5% 150mV VI
PW-IM110A-1Y240N 24V 4.49A 110W +/- 3% 150mV VI
PW-IM110A-1Y280N 28V 3.39A 110W +/- 3% 150mV VI
PW-IM110A-1Y360N 36V 3.06A 110W +/- 3% 150mV VI
PW-IM110A-1Y480N 48V 2.29A 110W +/- 3% 150mV VI
Note:
  • Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 47uF electrolytic capacitor and a
    0.1uF ceramic capacitor in parallel at output connector.