101W-250W Desktop Power Supply
210W AC-DC Desktop Adaptor

A08210-XX

210W AC-DC Desktop Adaptor
AC Inlet C14

Features
*Meet DOE Level VII and EU ErP Efficiency Compliant
*3-prong IEC320-C14, Class I AC inlet
*Medical safety approved ( 2x MOPP between primary to secondary )
Suitable for BF application with appropriate system consideration
*Altitude during operation : ITE up to 5000m , Medical Below 3000m
*< 0.15W No Load Input Power
*Approval IP42
*Touch current less than 100uA
*3 year warranty
View More Details
PDF Datasheet (562 KB)
Description
*Dimension : L 170×W 85×H 44 mm ( 6.69"x3.35"x1.73" )
*Weight : 0.78 kgs. ( 1.72 lbs. )
Specification
Electrical specifications
  • Input range :90 - 264VAC
  • Frequency :47 - 63Hz
  • Power Factor :>0.95 @115VAC; > 0.90 @230VAC @full load
  • Inrush current;< 25A peak @115VAC;< 50A peak @230VAC cold start @25C
  • Input current (rms) :2.5A @115VAC; 1.3A @230VAC max.
  • Efficiency :>92% @80% Full load, 230VAC
    >89% @Average efficiency, 115/230VAC
  • Maximum output power (Po) :210 Watts convection cooling
  • Hold-up time :>10ms typical @full load, 115VAC
  • Short circuit protection :Auto-recovery
  • Over load protection :Auto-recovery
  • Over voltage protection :Latch off
  • Over temperature protection :Latch off
Safety standards
  • UL/cUL 60601-1
  • TUV EN 60601-1
  • CB IEC 60601-1
  • CCC GB 4943. 1
  • UL/cUL UL 62368-1
  • TUV EN 62368-1
  • CB IEC 62368-1
EMC standards
  • EN 60601-1-2
  • EN 55011 Class B
  • EN 55032 Class B
  • EN 55035 Class B
  • CCC GB/T 9254. 1
  • CCC GB17625. 1
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature : -20 to +60°C (Refer to derating curve)
  • Operating Humidity: 10-95% , non-condensing.
  • Storage temperature : -20°C to +85°C, Non-condensing.
  • Storage humidity : 0-95% , non-condensing.
  • MTBF : > 350,000 hours @full load and 25°C ambient temperature based on Telcordia SR-332(Bellcore)
Model Output voltage Output current Po O/P Regulation Ripple & Noise
(Vp-p)
Efficiency
Level
A08210-12
+12V
17.50A
210W ±5% 100mV VII
A08210-15 +15V 14.00A 210W ±5% 100mV VII
A08210-19 +19V 11.05A 210W ±5% 100mV VII
A08210-24 +24V 8.75A 210W ±5% 150mV VII
A08210-28 +28V 7.50A 210W ±5% 150mV VII
A08210-36 +36V 5.83A 210W ±5% 150mV VII
A08210-48 +48V 4.37A 210W ±5% 150mV VII
Note:
  • Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 47uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector.
  • (-1 to -20°C ambient temperature and EMS Immunity worse case O/P Regulation≦ +/-10%)
  • The switching frequency of this series is set within 56 to 85KHz at full load.
  • The ripple and noise of this series is tested under full load condition.
  • Safety and EMC Performance
    Description Safety EMC
    Medical
    equipment
    IEC 60601-1:2005+AMD1:2012+AMD2:2020
    EN 60601-1:2006/A1:2013/A12:2014/A2:2021
    ANSI/AAMI ES60601-1:2005/A2:2021
    CAN/CSA C22.2 NO. 60601-1:14/A2:22
    EN 60601-1-2:2015+A1:2021
    EN 55011:2016+A1:2019+A2:2021
    FCC 47 CFR Part 18
    Audio / Video, ITEequipment IEC 62368-1:2018
    EN IEC 62368-1:2020+A11:2020
    UL 62368-1, 3rd Ed
    CAN/CSA C22.2 No. 62368-1:19, 3rd Ed
    EN 55032:2015+A11:2020
    EN 55035:2017+A11:2020
    FCC 47 CFR Part 15B(*)
    ICES-003 Issue 7
    Tests for conformance to this equirement will be performed with final system (*) FCC PART 15 compliance information and warings:
    This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions :
  • This device may not cause harmful interference, and
  • This device must accept any interference received, including interference that may cause undesired operation.

  • Insulation level and dielectric withstand (HI-POT)
    Audio / video, ITE equipment Isolation voltage Grade insulation
    Primary circuits to secondary circuits 4242Vdc (3000Vac) Reinforced
    Primary circuits to earth ground 2121Vdc (1500Vac) Basic
    Secondary circuits to earth ground 2121Vdc (1500Vac) Basic

    Medical equipment Isolation voltage Means of patient protection
    Primary circuits to secondary circuits 5656Vdc (4000Vac) 2 MOPP
    Primary circuits to earth ground 2121Vdc (1500Vac) 1 MOPP
    Secondary circuits to earth ground 2121Vdc (1500Vac) 1 MOPP
    Note : testing in Production use a dc voltage test 4 Sec.